Introduction:
In the first part of the project, the goal was to evaluate the maximum possible pressure resolution attainable using differential detection methods and a receiver utilizing spatial demultiplexing. A test procedure was outlined in the project proposal and QA narrative to determine the maximum resolution. In the previous report, difficulties in constructing the required experiments were reported. An optical spectrum analyzer (OSA) replaced the intended spatial demultiplexer in order to perform a preliminary evaluation of sensor resolution. Oil and water were successfully differentiated using a float sensor design. A diaphragm sensor design showed promise, but required an instrument with more resolution than the OSA.
Since the last report, the spatial demultiplexing receiver was successfully implemented and thoroughly tested. This report describes the results of these tests and the procedures used. The receiver is capable of higher wavelength resolution (and therefore pressure resolution) than the OSA.
Receiver
Optimization:
The
receiver configuration under evaluation is depicted in Figure 1. As mentioned in the last report, we are
using a grating with less-than-optimal properties due to difficulties in
securing an ideal grating from a manufacturer.
Since last report, an imaging lens has been added. The lens lightly focuses the incoming light
onto the camera. The location of the
focused light depends on the location of the incident light on the lens, which
is in turn dependent on the wavelength.
As mentioned in the QA narrative, there exists a design trade-off
between the size of the focal spot and the instrument resolution.
Our efforts focused on the
proper placement of the lens with respect to the grating and the camera, and
the proper choice of the lens’ focal length.
Three different lenses having focal lengths of 19mm, 38mm, and 50mm were
evaluated. Each lens was positioned at
several distances d from the camera,
starting with d equal to the focal
length and increasing in 10% increments.
The distance was further refined by identifying the two distances that
provided the best resolution and positioning the lens at a point halfway
between these distances. The lens
position with respect to the grating was determined with multiple sensor
signals, as the lens diameter limits the maximum spatial difference between any
two wavelength to be detected.
Results Since Last Report:
The first experiments
focused on selecting the proper focal length of the lens and positioning with
respect to the camera. For the
experiments, a single sensor signal was used, and the signal wavelength varied
using a piezo-electric stretcher.
Wavelength changes were calibrated using the OSA at selected values of
the voltage applied to the stretcher.
The lens position was adjusted so that the focused light formed a
circular spot on a relatively noise-free part of the camera. A circular spot ensures that the incoming
optical rays are parallel to the optical axis of the lens, and therefore
changes in location and shape of the recorded light pattern are due only to
wavelength changes. Camera data was
converted to a bit-map file and then into spreadsheet format for performing
data processing. The centroid algorithm
described in the QA narrative was applied to the data using MS Excel. Before applying the algorithm, however, some
pre-processing was required, as light levels often caused the digital output
from the camera to “cycle” one or more times – that is to reach 255 and cycle
back to 0 one or more times for high optical powers incident at the pixel
location.
Typical results as a
function of focal length are shown in Figure 2. The 50mm focal length lens did not produce significant movement
in the spot location on the camera for most distances, and was quickly
discounted. The 38mm lens produced
significant movement (c.f. Figure 3), with an optimal location at 1.36 times
the focal length, as determined by measurements of displacement and standard
deviation at each position. For a 100V
volt change on the piezo-electric controller, corresponding to a 0.26nm change
in wavelength, a 4.9 pixel change in location was achieved. Thus the linear
dispersion at the camera input was 0.053 nm/pixel. As before, the minimum resolution of the instrument was set by
twice the standard deviation of the measurements. The minimum resolution
afforded by the 38mm lens was 0.0387 pixels, corresponding to 2.05 pm. The 19mm lens produced an even higher
resolution of 0.060 nm/pixel, but at a distance of 1.2 times the focal
length. Typical results from the 19mm
lens are shown in Figure 4.
A second sensor was added to
the test configuration to determine whether the lens was capable of capturing
both wavelengths simultaneously at the distance for optimal resolution. The wavelengths of the two sensors were
separated by 46.085 pixels when both sensors were at rest. A one-inch diameter lens was just able to
collect both wavelengths without significant diffraction of light from the
edges of the lens, which would increase system noise. No measurable increase in diffraction was recorded when 17.2 V
was applied to the piezo-electric driver on either sensor.
For the current experimental
parameters, a maximum of four sensors per fiber string could be discriminated
accurately. The primary limitation on
system performance is the still the grating we are currently using. An imaging (concave) grating or one with a
greater linear dispersion would improve both the resolution and the range of
wavelengths that could be collected.
In the two sensor
configuration, an experiment was conducted to determine the improvement in
resolution of the receiver using differential measurement. The experiment was performed using the 38mm
lens. Results of the experiment are
shown in Table 1 and Figures 5 and 6.
On average, the resolution was improved by a factor of 44.3%, from
0.0595 to 0.0362 pixels. Based on these
results, we expect that water and oil can be differentiated with the
appropriate sensor design.
We are scheduled to begin
resolution tests within the oil-water column in the next several weeks. As noted above, the resolutions obtained
with the 38mm and 19mm lenses indicates that we will be able to differentiate
between oil and water for both sensor designs.
A diaphragm-based design is more desirable at present, since we believe
it can be implemented with a minimum of moving parts and without exposure of
the fiber optic cable to the external (and often hostile) environment present
in a production column. We are
conferring with the project consultants on the sensor designs and practical
concerns in implementing the receiver.
Mirror GRATING Fiber Output

Figure 1: Extended path experimental setup for spatial de-multiplexing
receiver.

Figure 2: Displacement at camera plane as a function of the
lens focal length. Displacement is
measured with respect to the voltage change on the piezoelectric controller.

Figure 3: Shift in the peak location on the camera as a function of the
distance between lens and camera for the 38mm focal length lens.

Figure 4: Peak location in pixels as a
function of the voltage setting of the piezoelectric controller.

Figure 5: Typical
processed output for dual sensor measurement.
The pixel value has been adjusted for noise and for rollover in the
quantized digital output.

Figure 6: Results of
differential measurement experiment.
Voltage and Position difference are sensor 2 – sensor 1. Note that there are two sets of measurements
at –8.6V – one for 0V – 8.6V, and one for 8.6V – 17.2V.
Table 1: Summary of resolution data for dilfferential measurement. Measurements are quoted in terms of standard
deviation.
|
[V1, V2] |
[0, 0] |
[0, 8.6] |
[17.2, 8.6] |
[0, 17.2] |
[8.6, 0] |
|
s1
+ s2 (pixels) |
0.0392157 |
0.0473458 |
0.0406284 |
0.0833363 |
0.0869257 |
|
s(2-1) (pixels) |
0.0233435 |
0.0177992 |
0.0130581 |
0.0658063 |
0.0607721 |
|
% change |
-40.5 |
-62 |
-67.9 |
-21 |
-30 |